A Synspec GC is the ideal choice for companies, institutions or authorities that want to measure with a gas chromatograph.
The Synspec GC has not been developed for analysis in the laboratory, but especially for continuous use under difficult conditions, e.g. in immission measuring stations.
A special training as chemist is not necessary, a simple instruction is sufficient.
The sampling system is equipped with a stepper motor controlled sampling piston, which enables very accurate and reproducible volume sampling. This also allows an automatic linearization of the system with a test gas concentration. This is based on the possibility to automatically reduce the sampling volume by means of the very precise stepper motor control (40ml with 1500 motor steps).
For this series, various sampling systems (loop, enrichment), column circuits (stripper, backflush, parking circuits) with one or two multi-way valves, detectors (PID, FID, WLD, ECD), double detector systems and double analytical systems are available.
GC 955 Model 601 PID Certified to DIN EN 15267 QAL1 2018-04-23
Some examples of standard applications of the GC955:
Workplace measurement of toxic components: e.g. for benzene, 1,3 butadiene, vinyl chloride, ethylene oxide, propylene oxide, epichlorohydrin, methyl isocyanate, captans, etc.
Series 600 BTX analyzer for ambient air applications: e.g. for benzene, toluene, ethylbenzene, m-xylene, p-xylene, o-xylene
POCP / VOC for measurement of ozone precursors contains over 30 pre-calibrated hydrocarbons in the range C2 - C10